Description:
Operating principle: magnetic induction/eddy current (F/NF)
With separate probe(s)
Measuring range: 0-1250um/0-50mil
Resolution: 0.1/1
Accuracy: ¡À1-3%n or ¡À2.5um
Min. measuring area: 5mm
Min. sample thickness: 0.3mm
Battery indicator: low battery indicator
Metric/ imperial: convertible
Power supply: 4x1.5V AAA(UM-4)battery
Auto power off
Operating conditions: 0-+45¡æ(32¨H-104¨H),¡Ü90%RH
Dimensions: 124x62x30mm
weight: 115g
Standard Delivery of TG8829FN Main unit 1 F type probe 1 N type probe 1 F calibration base set 1 N calibration base set 1 Calibration foil set 1set (4 foils) Carrying case 1 Instruction manual 1
Optional Accessories Cable and software for RS-232C
Customized testing range
|